BASIC METHODS FOR HANDLING & CARE OF INSTRUMENTS CLEANING, STERILIZATION & MAINTENANCE OF BEAUTY INSTRUMENTS

New instruments should be cleaned, dried and sterilized before the first use.

1.CAGE IN /PRE-BATHE

it is important never to hold instruments in a dry container, which allows blood and waste to dry onto instrument surfaces and makes cleaning more difficult. If rinsing and decontamination processes are not immediately available, pre-treat instruments or hold in a neutral holding/pre-bathe enzymatic solution after patient use but before actual cleaning. Prep Enzyme Foaming spray for pre-cleaning of dirty instruments and scopes. Simply spray on instruments until ready for rinsing and disinfection. Also, use neutral pH all purpose multi-enzyme concentrate is ideal for instrument presoaking and pre-cleaning. As soon as possible, rinse, disinfect and clean as follows:

 

2.RINSING

Immediately, remove organic materials by rinsing instruments under warm (not hot) running water after surgery. Rinse should remove most blood fluids and tissue. Do not process metals (stainless, copper, chrome plated, etc.) together. Always wear safety protection gear.

3. DISINFECTING

To protect medical personnel from contamination during cleaning, immerse instruments completely in an EPA approved disinfectant for approximately 10-20 minutes. Always closely follow manufacturers’ recommended disinfecting time and solution preparation and instruction. Then rinse again.

CAUTION

Disinfected instruments are NDT STERILE. Never expose stainless steel instru­ments to bleach or other corrosive chemicals to disinfect. Exposure to bleach may result in instruments pitting and will cancel all guarantees by the manufac­turer.

4. CLEANING

All blood dried body fluids and should be completely removed from the instru­ments prior to sterilization. Several methods are available.

A. BATHE

An enzymatic cleaner bath (soak) or a solution of water and neutral pH (7) detergent are effective in removing organic material from instruments. Use distilled (demineralized) water if possible. Instruments should be fully submerged for at least 1 O minutes.Do not let “sharp• (scissors, knives, osteo­tomes, etc.) running water to remove solution. Change solutions frequently.

B. ULTRASONIC CLEANING

Most instruments manufacturers recommend ultrasonic cleaning as the most effective way to clean surgical instruments, particularly those with hinges, box locks and other moving parts. All instruments must be fully submerged in open position. Use distilled (dernineralized) water 1f possible. Make sure that ‘sharp• blades do not touch other instruments to prevent possible surface scratching and also be sure dissimilar metal instru­ments are separated. Process instruments for full recommended ultrasonic cleaning cycle. Change solution frequently or as often as the manufacturers recommends. Rinse instruments with water to remove the cleaning solution, can be safety use with all ultrasonic unites.

C. AUTOMATIC WASHER STERILIZER

Follow manufacturers’ recommen­dations but ensure instruments are lubricated after the last rinse cycle and before the sterilization cycle.

 

CAUTION

• Needle holder and forceps may check if sterilized with ratchet in closed position. Manual Cleaning: if ultrasonic cleaning is not available, observe the following steps.
• Use stiff nylon cleaning brushes. Do not use steel wool or wire brushes except specially recommended stainless steel wire brushes for instruments serrated areas . bone or on stained areas of knurled handles. 

. . Use only neutral pH (7) detergents such. If not rinsed off property, low pH  (ac1d1c – less than 6 pH) detergents break down the stainless protective surface result­ing in pitting and/or back staining. High pH detergents (alkaline – more than 8pH) which can also interfere with the smooth operation of instruments. Most brown stains are not rust and are easily removed with stain remover.

• Brush delicate instruments carefully, and if possible, separate them from general instru­ments.  closed smoothly. Test cutting performance at 34 length of the blade with the

following recommended materials. Scissors should cut all the tips. Recom­mended cutting test materials:
Fine/Delicate scissors: Surgical gloves
, Medium scissors: Single layer of stocking/cast netting
, Large/utility scissors: Double layer of stocking/cast netting.
, Check forceps (pickups) for proper jaw alignment. Teeth must meet properly –
without catching.
, Check hemostats and needle holders to ensure jaw tips close in first ratchet position and entire jaw should close in third ratchet position. Check instruments for loose hinges and verify that they lock and unlock easily.
 Suction tube should be clean inside.
Test Biopsy Punches by punching a clean hole in 3-6 mil thick poly-bag material. If poly-bag material is not available, use tissue paper.
,Retractors should function properly.
, CUiiing edge instruments and knives should be sharp and free of damage. . , After scrubbing, rinse instruments thoroughly under running water. While rinsing, open and close scissors, hemostats, needle holder and other hinged instruments to make sure the hinge areas are rinsed out and no waste remains.

5. AFTER CLEANING

Separate dissimilar metals prior to sterilizing/autoclaving. If instruments are to be stored, let them air-dry and store them in a clean and dry environment.

6. AUTOCLAVING

A. Lubricate all hinged instruments which have any “metal to metal” action, at the screw or box lock.
A non-silicone, water-solubles B. Sterilizer instruments either individually or in sets. Individual Instruments: Disposable paper or plastic pouches are ideal. Make sure to use a wide enough pouch for instruments with ratchet locks so instruments can be sterilized in an open (unlocked) position. Instruments locked during auto­claving can experience cracked hinges (box lock) or other problems because of heat expans ion. If wrapping instruments, make sure the towel does not contain detergent residue, which can stain instruments.
Instruments Sets: Unlock all instruments and sterilize in an open position. Place heavy instruments on the bottom of sets (when two layers are required). Do not overload the chamber because an air pocket may form that hinders steam penetration. 

CAUTION

With most portable tabletop autoclaves, at the end of the autoclave cycle and before the drying cycle, unlock the door and open it no more than crack, about¼” (6.4mm). Then run the dry cycle for the period recommended by .the autoclave manufacturer. If the autoclave door is fully opened before the drymg cycle, cold room air will rush into the chamber causing condensation on the instruments that may result in water stains or cause wet packs. Make sure autoclave filters and chambers are cleaned as recommended by the manufacturer. Stain Remover is effective is removing stains and cleaning autoclave chamber.

7. CHEMICAL /COLD STERILIZATION

Most chemicals/cold sterilization solution renders instruments sterile only after 10- hour immersion. This prolonged chemical action can be more detrimental than usual 20-minute autoclave cycle. If the instruments need to be “disinfected” only, a chemical/cold sterilization soak is acceptable, as disinfection in using bleach (paragraph 3). Sterile-an absolute term (no living Organism survives) Disinfected-basically clean. Some organisms may survive. Always use the proper sterilization/cleaning Technique to render he instruments in the required condition for use.

 

CAUTION

For instruments with tungsten carbide inserts jaws, we do not recommended use of chemicals/cold sterilization solution, which may deteriorate the instrument’s jaw. Delicate instruments and Mouth Mirrors should be cleaned by hand. Do not use ultrasonic cleaning machines